Channel A can
compensate up to 10 pF of electrode shunt capacity and Channel B can
compensate up to 50 pF.
Tickler
Circuit
Assists in
cell penetration. The frequency and amplitude of the oscillations may be
varied for differences in membrane thickness or cell size. The duration of
tickle can be controlled either by using the momentary switch, a foot
switch, or by applying a signal to the remote tickler input.
Active
Filters
Low pass
settings on a -40 dB/decade active filter vary the cutoff from 1 to 30
kHz. Either probe or bridge outputs may be selected for filtering.
Current
Injection
Channel B can
eject current through the microelectrode by applying a command signal to
the stimulus input connector; the resulting output from the probe will
then be a constant current replica of the input signal. Two ranges of
current delivery are provided: 50 nA and 500 nA or by an external source.
This source can be useful for delivering hyperpolarizing currents to
stabilize the cell membrane potential and as a holding current for
microiontophoresis.
Bridge
Balance
Subtracts the
excess electrode voltage associated with delivering current through the
recording micropipette. Electrode resistances up to 1000 M£[ can be
balanced in two ranges. The balanced signal is available from x10 or x50
front panel output connectors.
Independent
Outputs
HEADSTAGE (PROBE) |
712P (red, port B), 715P (blue, port A) |
ACTIVE PROBE INPUT IMPEDANCE 712P (red,
port B) |
>1011ohm |
ACTIVE PROBE INPUT IMPEDANCE 715P (blue,
port A) |
1015ohm |
GAIN 712P (red, port B) |
x1, x10 |
GAIN 715P (blue, port A) |
x1 |
OUTPUT RESISTANCE 712P (red, port B) |
100ohm |
OUTPUT RESISTANCE (blue, port A) |
100ohm |
OUTPUT VOLTAGE RANGE 712P (red, port B) |
¡Ó10 V |
OUTPUT VOLTAGE RANGE (blue, port A) |
¡Ó 10V |
MAXIMUM INPUT VOLTAGE 712P (red, port B) |
¡Ó15V |
MAXIMUM INPUT VOLTAGE (blue, port A) |
¡Ó15 V |
PROBE LEAKAGE CURRENT 712P (red, port B) |
5 X 10-12 A |
PROBE LEAKAGE CURRENT (blue, port A) |
10-14 A |
DC POSITION ADJUST RANGE 712P (red, port
B) |
¡Ó 300 mV |
DC POSITION ADJUST RANGE (blue, port A) |
¡Ó 300 mV |
ELECTRODE RESISTANCE TEST CURRENT 712P
(red, port B) |
1nA |
ELECTRODE RESISTANCE TEST CURRENT (blue,
port A) |
1pA, 1nA selectable |
INPUT CAPACITY COMPENSATION 712P (red,
port B) |
+10 to -50Pf |
INPUT CAPACITY COMPENSATION (blue, port
A) |
0 to -10pF |
NOISE:Input shorted 712P (red, port B) |
|
NOISE:Input shorted (blue, port A) |
|
NOISE:20 MΩ carbon resistor
712P (red, port B) |
|
NOISE:20 MΩ carbon resistor
(blue, port A) |
|
RISE TIME:10-90% direct input small
signal |
1us, typical |
RISE TIME:10-90% through 20MOhm (-C
"on") |
25us, typical |
CURRENT INJECTION: Internal DC Current
712P (red, port B) |
¡Ó 50nA low range, ¡Ó 500nA high range |
CURRENT INJECTION: Externally commanded
Current 712P (red, port B) |
¡Ó 500nA low range, ¡Ó5uA high range |
CURRENT INJECTION: External current
command factor 712P (red, port B) |
20mV/nA low range, 2mV/nA high range |
CURRENT INJECTION: Current monitor 712P
(red, port B) |
100mV/nA low range, 10mV/nA high range |
CURRENT INJECTION: Compliance 712P (red,
port B) |
3V low range, 10V high range |
CURRENT INJECTION: Bridge balance 712P
(red, port B) |
0-100ohm, 0-1000ohm |
CURRENT INJECTION: Bridge amplifier gain
712P (red, port â£áœBâ£á?) |
x 10, x 50 |
LOW PASS FILTER |
40 dB/decade, continuously variable 1-30
kHz |
METER SECTION: Display |
3.5-digit LED |
METER SECTION: Ranges |
200 mV, 2000 mV, 20 V, 200 nA, 2000 nA |
METER SECTION: Accuracy and resolution |
1 digit |
DIMENSIONS: Instrument |
17 x 5.25 x 10 in. (43 x 13 x 25 cm) |
DIMENSIONS: Probe |
Diameter: 12 mm Length: 34mm |
POWER |
95-135V or 220-240V, 50/60Hz |
SHIPPING WEIGHT |
15 lb (7 kg) |
CERTIFICATION |
CE, CSA |